Operations 3.0 – Teledyne 275R - Portable turbine generator purge gas analyzer User Manual

Page 11

Advertising
background image

3.3

TELEDYNE ANALYTICAL INSTRUMENTS

Operations 3.0

Operations 3.0

Operations 3.0

Operations 3.0

Operations 3.0

2. With a mixture of 90% H

2

and 10% N

2

flowing through the

analyzer, adjust the zero control potentiometer to read zero on
the meter for Range 3.

3. With 100% H

2

flowing through the cell, adjust the span controls

to read full scale output in Ranges 1 and 3.

4. With air flowing through the cell, adjust the span control to read

full scale output in Range 2.

After calibration has been successfully concluded, connect the sample

gas to the analyzer, and adjust the flow to approximately 0.1 SCFH

0.1 SCFH

0.1 SCFH

0.1 SCFH

0.1 SCFH.