Teledyne 311 - Portable trace oxygen analyzer User Manual
Page 16

7.
PRODUCT SPECIFICATION DATA
Ranges:
0-10, 0-100, 0-1000, 0-10000 ppm Oxygen
CAL range for air calibration
Sensitivity: 0.5% of Full Scale
Accuracy:
±2% of full scale (except ±1 ppm for 0-10 ppm range) at constant
temperature and pressure (temperature and pressure of calibration).
±5% of full scale (except ±1 ppm for 0-10 ppm range), over
operating temperature range (once temperature equilibrium has been
achieved).
System Operating Temperature: 0-50
° C
Response Time:
90% response time 61 seconds
Reproducibility:
±1% at constant temperature
Sensor Type: Class
B-2C
System Power Requirements:
Model 311 - AC power for battery recharge circuit of two
current limited rechargeable NiCad batteries, 115 VAC,
50/60Hz
(100/220
VAC
optional).
Weight:
6 lb (2.71 Kg.).
Approval:
Intrinsically Safe (Class I, DIV. I, GROUP A, B, C and D), Factory
Mutual
approved.
-13-