1-6

1 Introduction

Model 3000MA

Teledyne Analytical Instruments

•

Alarm Connections

2 concentration alarms and 1 system
alarm.

•

RS-232 Port

Serial digital concentration signal output
and control input.

•

Remote Probe

Used in the 3000MA for controlling
external solenoid valves only.

•

Remote Span/Zero

Digital inputs allow external control of
analyzer calibration. (See Note, below.)

•

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.

•

Range ID Contacts

Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.

•

Network

For future expansion. Not implemented
at this printing.

Optional:

•

Calibration Gas Ports

Separate fittings for zero, span and
sample gas input, and internal valves
for automatically switching the gases.

• Current Signal Output Additional isolated 4-20 mA dc plus

4-20 mA dc range ID.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 3000MA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.