5 Maintenance

Model 3000PB

5-8

Teledyne Analytical Instruments

During preamp testing there is a countdown in the lower right corner

of the screen. When the testing is complete, the results are displayed.

Power: OK Analog: OK
Preamp: 2

The module is functioning properly if it is followed by OK. A number

indicates a problem in a specific area of the instrument. Refer to Table 5-1
for number-code information. The results screen alternates for a time with:

Press Any Key
To Continue...

The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed

Analog

0

OK

1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3

Both Failed

Preamp

0

OK

1

Zero too high

2

Amplifier output doesn't match test input

3

Both Failed