1-5
Trace Oxygen Analyzer
Introduction 1
Teledyne Analytical Instruments
1.7
Rear Panel (Equipment Interface)
The rear panel, shown in Figure 1-2, contains the gas and electrical
connectors for external inlets and outlets. Some of those depicted are op-
tional and may not appear on your instrument. The connectors are described
briefly here and in detail in chapter 3 Installation.
Figure 1-2: Model 3000TA Rear Panel
•
Power Connection
Universal AC power source.
•
Gas Inlet and Outlet
One inlet (must be externally valved)
and one exhaust out.
•
Analog Outputs
0–1 V dc concentration plus 0-1 V dc
range ID, and isolated 4–20 mA dc plus
4-20 mA dc range ID.
•
Alarm Connections
2 concentration alarms and 1 system
alarm.
•
RS-232 Port
Serial digital concentration signal output
and control input.
•
Remote Probe
Used in the 3000TA for controlling
external solenoid valves only.
•
Remote Span/Zero
Digital inputs allow external control of
analyzer calibration.