Teledyne 3000TA-EU - General purpose trace oxygen analyzer User Manual

Page 48

Advertising
background image

Operation 3000TA-

EU

Teledyne

Analytical

Instruments

36

4.3.2 Setting up an Auto-Cal

When proper automatic valving is connected (see Chapter 3,

Installation), the analyzer can cycle itself through a sequence of steps
that automatically zero and span the instrument.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model 3000TA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.

To setup an Auto–Cal cycle:
Choose System from the Function buttons. The LCD will display

five subfunctions.

TRAK/HLD Auto—Cal
PSWD Logout More

Use ◄►arrows to blink

Auto—Cal

, and press Enter. A new screen

for Span/Zero set appears.

Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h


Press ◄►arrows to blink

Span

(or

Zero

), then press Enter again.

(You won’t be able to set OFF to ON if a zero interval is entered.) A
Span Every ... (or Zero Every ...) screen appears.

Span Every 0 d
Start 0 h from now


Use ▲/▼ arrows to set an interval value, then use ◄ arrows to

move to the start-time value. Use ▲/▼ arrows to set a start-time value.

To turn ON the Span and/or Zero cycles (to activate Auto-Cal):

Press System again, choose

Auto—Cal,

and press Enter again. When

the Span/ Zero values screen appears, use the ◄► arrows to blink the
Span (or Zero) OFF/ON field. Use

▲/▼ arrows to set the OFF/ON

field to ON. You can now turn these fields ON because there is a
nonzero span interval defined.