Appendix – Teledyne 3000TA-XLEU - Trace oxygen analyzer User Manual
Page 99

Trace Oxygen Analyzer
Appendix
Teledyne
Analytical
Instruments
83
Appendix
A-1 Specifications
Packaging: General
Purpose
• Flush panel mount (Standard)
• Relay rack mount. Contains either one
or two instruments in one 19" relay rack
mountable plate (Optional).
Sensor: Teledyne B-2CXL trace analysis Micro-
fuel Cell.
Insta-Trace Micro-fuel Cell (option)
Cell Block: 316 stainless steel.
Sample System: All wetted parts of 316 stainless steel.
90 % Response Time: 65 seconds at 25°C (77°F) on 10 ppm
range.
100 seconds on 1 ppm range.
Ranges: Three user definable ranges from 0–1 ppm
to 0–250,000 ppm, plus air calibration
range of 0-250,000 ppm (25%).
Autoranging with range ID output.
Alarms: One system-failure alarm contact to detect
power failure or sensor zero failure.
Two adjustable concentration threshold
alarm contacts with fully programmable
setpoints.
Displays: 2-line by 20-character, VFD screen, and
one 5 digit LED display.
Digital Interface: Full duplex RS-232 communications port.