1 Introduction

Model Ultra Trace 3000

Teledyne Analytical Instruments

•

RS-232 Port

Serial digital concentration signal output
and control input.

•

Remote Probe

Used in the Ultra Trace3000 for
controlling external solenoid valves
only.

•

Remote Span/Zero

Digital inputs allow external control of
analyzer calibration.

•

Calibration Contact

To notify external equipment that
instrument is being calibrated and
readings are not monitoring sample.

•

Range ID Contacts

Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.

•

Network I/O

Serial digital communications for local
network access. For future expansion.
Not implemented at this printing.

Note: If you require highly accurate Auto-Cal timing, use external

Auto-Cal control where possible. The internal clock in the
Model Ultra Trace 3000 is accurate to 2-3 %. Accordingly,
internally scheduled calibrations can vary 2-3 % per day.