Appendix – Teledyne 3010MA - Paramagnetic oxygen analyzer, includes 0-100% range User Manual
Page 81

Paramagnetic O2 Purity Analyzer
Appendix
Teledyne Analytical Instruments
81
Appendix
A-1 Model 3010MA Specifications
Sensor:
Paramagnetic
Range:
3 ranges, user selectable from 0-100% O
2
Alarms:
One system-failure alarm contact to detect
power failure
Two adjustable concentration threshold
alarms with fully programmable setpoints.
Displays:
2 line by 20 character, alphanumeric, VFD
screen
One 5 digit LED display
Digital Interface:
Full duplex RS-232 communications port.
Power:
100-130, or 200-240 VAC, at 47/63 Hz.
(User selected on rear of control unit)
Operating Temperature: 20-30
°C
Accuracy:
±0.045%
O
2
absolute at constant
temperature (once thermal equilibrium has
been achieved)
±0.05% O
2
over operating temperature range
20-30
C (once thermal equilibrium has been
achieved)
Response
Time:
10% - 90% in 14 sec @ 600 SCCM
Analog outputs:
0-1VDC percent-of-range (Standard)
0-1VDC range ID (Standard)
4-20 mA dc—isolated— percent-of-range
4-20 mA dc—isolated— range ID (Special
option)
Dimensions:
17.7 cm high
24.9 cm wide 30.5 cm
deep (6.96
8.7 12.2).
Weight:
Analysis:
(7.8Kg)
17.2Lbs
Control
unit:
(8.2 Kg) 18Lbs