Teledyne 3220 - Multi-channel oxygen monitor system User Manual
Page 34

3 Installation
Model 3220
Teledyne Analytical Instruments
20
3.6.2 Control Unit Jumper Settings
The Control Unit may be configured by changing the installation of
the jumpers on the PCB. See the following sections for complete
jumper installation information. The default jumper installation is
shown in the table below.
Table 3-1: Control Unit Jumper Configuration
Jumper Installation Function
JP1
Pins 1 & 2
Fail-safe Failure Alarm
JP2
Pins 1 & 2
Fail-safe Alarm 2
JP3
Pins 1 & 2
Fail-safe Alarm 1
JP4
Pins 1 & 2
Normally closed ext.
aud. Alarm
JP5
Pins 1 & 2
Fail-safe ext. aud.
Alarm
JP6
Pins 1 & 2
Ext. aud. contact
closures output
JP7
Pins 1 & 2
Ext. aud. contact
closures output
JP8 Thru JP16
OFF
RS-485 connections
not routed
JP17
ON
Internal aud. Alarm
enabled
JP18 & JP19
ON
Ground planes
connected
Three alarms (Alarm 1, Alarm 2, and Fail) may be configured for
either fail-safe (default) or non fail-safe operation according to the table
below:
Table 3-2: Fail-safe/Non Fail-safe Alarm Configuration
Alarm Jumper Failsafe
Non-failsafe
Alarm 1
JP3
Pins 1 & 2
Pins 2 & 3
Alarm 2
JP2
Pins 1 & 2
Pins 2 & 3
Failure
JP1
Pins 1 & 2
Pins 2 & 3