Teledyne 402REU - Hydrocarbon analyzer User Manual
Page 53

A-1
Teledyne Analytical Instruments
Total Hydrocarbon Analyzer
Appendix
IF USED OUTSIDE OF THESE SPECIFICATIONS
THE INSTRUMENT MAY BE DAMAGED AND
HAZARDOUS CONDITIONS MAY BE PRODUCED.
Appendix
Specifications
Ranges: 0-10, 0-100, 0-1000ppm hydrocarbon stan-
dard (methane equivalent)
(0-1ppm low range is optional.)
Sensitivity: 1% of full scale
Accuracy: ±2% of full scale at constant temperature
(Applies to standard ranges.)
Response Time: 90% in less than 15 second
Operating Temperature: 41°F to 110°F (5°C to 43.3°C)
Fuel: Mixture of 40% hydrogen, 60% nitrogen
recommended
(Other mixtures are possible. Contact TET/AI
for applications assistance.)
Sensor Type: Flame Ionization Detector with Flame Guard
Signal Output: Internal - Digital Panel Meter
External - 0-1VDC, 4-20mA DC Isolated
Power Requirement: 115VAC ± 10% 50/60 Hz
230VAC ± 10% 50/60Hz
90VA nominal, 150VA maximum
Enclosure: 19" (48.3cm) Rack-Mounted Steel Enclosure
19" (48.3cm) W x 8.75" (22.2cm) H x 15.5"
(39.4cm) D