Operation

Teledyne Analytical Instruments 51

4.4.14 Analog output Adjustment

The software in this instrument provides a way to enter to manage

the analog output of the analyzer FOR ANALOG CHANNEL 1
ONLY
. To access the offset function:
<▲▼>

From the MAIN MENU, scroll to the ANALOG
ADJUST function using the UP/DOWN keys.

<Enter>

Pressing <Enter> activates the function and takes you to
the next screen.

ANALOG-OUT

ADJUST

-----------------------

CH1 Offset: 0

CH1 Gain: 0

CH1 Select: THC

CH2 Select: R-ID

-----------------------

<Enter>

Use the <Enter> key to move over to the Offset field.

<▲▼>

Use the UP/DOWN keys to change each digits value.

<Enter>

Use the <Enter> key to save the value.

<Escape>

Use the <Escape> key to revert to the original value.

Use the same procedure to set the gain of the instrument. After the

last digit is entered, the final <Enter> press will accept the gain value.

The values entered do not have explicit units of current or voltage.

Adjustment should be gauged using a voltmeter or current meter
attached to the analog output. When the ANALOG ADJUST menu is
entered the following is displayed:
________________________________________________

-> CH1 mA Offset: 0

CH1 mA Gain: 0

________________________________________________

When the arrow is on the Offset, or when the user presses <Enter> to

modify the offset, the analog output is 4mA plus a value that is
proportionate to the number indicated to the right of 'Offset:' That
number may be modified –500 to 500.

When the arrow is on the Gain, or when the user presses <Enter> to

modify the gain, the output is 20mA scaled by the value to the right of