Teledyne 4020 - Hydrocarbon analyzer User Manual
Page 39

Total Hydrocarbon Analyzer
Installation
Teledyne
Analytical
Instruments
23
When CRC opens again, zero and span are done, and the sample is
being analyzed.
Note: The remote valve connections (described below) provides
signals to ensure that the zero and span gas valves will be
controlled synchronously.
3.3.3.7
R
ANGE
ID
R
ELAYS
There are three dedicated Range ID relay contacts. They are
assigned to relays in ascending order—Low range is assigned to Range
1 ID, Medium range is assigned to Range 2 ID, and High range is
assigned to Range 3 ID. Table 3-6 lists the pin connections.
Table 3-6: Range ID Relay Connections
Pin
Function
21
Range 1 ID Contact
38
Range 1 ID Contact
22
Range 2 ID Contact
39
Range 2 ID Contact
19
Range 3 ID Contact
18
Range 3 ID Contact
34
Not
Used
35
Not
Used
3.3.3.8
N
ETWORK
I/O
A serial digital input/output for local network protocol. At this
printing, this port is not yet functional. It is to be used in future options
to the instrument. Pins 13 (+) and 29 (–).
3.3.3.9
P
IN
O
UT
T
ABLE
The following table summarizes all the outputs/inputs available in
the 50 pin D-Sub connector on the backpanel of the analyzer.