Teledyne 6650 - Fluorescence Probe - Oil in Water analysis system User Manual
Page 35

UV-Photo-X Specifications
Teledyne
Analytical
Instruments
27
6.2 Transmitter
Specifications
1) Configuration.XSource Reference Through Media
Reference.
Application specific settings
Operation mode Display format XXXX
specified 4mA = (unspec, field settable)
specified 20mA = (unspec, field settable)
2) Optical specifications.
a) Lamp.
Xenon Flash Lamp, Life: 3-5 years
b) Detectors.
PMT
Spectral range: UV 190-690 nm
3) Electrical specifications. ( Tungsten Halogen, X Flash
Lamp)
a) Power Requirements 24VDC(IEC connector)
Current Requirements; LED ~332mA Tungsten
Halogen ~728.5mAmA (24VDC)
Consumption:
LED 8 W Tungsten Halogen ~17.5W (24VDC)
b) Output signal: Isolated 4-20mA sourced output
Permissible loop resistance
500 ohms for customer use
c) Span calibration:
Remote trigger via normally closed contact (must remove the
jumper on the top of the Photo-X unit for remote triggering)