Welch Vacuum 2585B-50 WOB-L Pinston Pump User Manual

Page 9

Advertising
background image

9

Section 3: OPERATION


3.10 Starting

Procedures

3.11 Starting a Welch Dry Vacuum Pump

Before attaching the pump to a system it is well to familiarize yourself with the function and action of

the pressure vacuum pump that you have acquired. Review the power requirements as described in

Section 2.6. Welch recommends running the pump for a few minutes to warm up before use. The

warm-up improves the pump’s ability to handle humid air.

3.12 Cleanliness

Take every precaution to prevent foreign particulates from entering the pump. Particulates will damage

the pump’s performance. If you fi nd that particulates will come off the process during evacuation, a

particle trap in the foreline will work. A simple, inexpensive trap may be made by placing glass wool in

a glass or plastic tube. Screens must be inserted to hold the glass wool in place.

3.13 Leak

Detection

The importance of eliminating all leaks in a vacuum system is obvious. The pump must remove this

added volume of leaked gas to maintain the desired vacuum. Leaks for these pumps can be located by

slightly pressuring the system and painting the suspected area with a thick soap solution. Escaping air

will produce soap bubbles.

3.14 Operating Pressure Range

Model 2585 Dry Vacuum Pumps are designed to be run from atmospheric to their maximum vacuum

level on the intake side. Consult the Specifi cation Table in the back of this manual for the ratings for

your

specifi c model.

3.15 Shutdown

Procedures

After use, Welch recommends the pump be run for about 2 minutes disconnected from the vacuum

process. The air pumped through the mechanism will purge out water vapor or droplets of water

condensate that may have formed on the inside of the pump. This purge of the pump mechanism helps

prevent

corrosion.

Section 4: MAINTENANCE


Welch dry pressure\vacuum units are 100% oil-free. The pump employs a non-lube piston and cylinder.

No maintenance is necessary for the bearings. All bearings are sealed and permanently lubricated.

Lubrication should not be attempted. The units are built for continuous duty operation with the

quietness and durability of a diaphragm, but with piston pump performance.

WARNING

The Pump is not recommended for pumping acid, base or organic

vapors or gases. Serious damage to the pump will shorten the

pump’s service life.In addition, pumping fl ammable vapors or

gases can lead to serious safety hazard leading to fi re or explosion.