Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide

215

Creating and Downloading Waveform Files

Creating Waveform Data

2. Save the I/Q data to a text file to review.

The following lines of code export the I and Q data to a text file for validation. After exporting the
data, open the file using Microsoft Excel or a similar spreadsheet program, and verify that the I and
Q data are correct.

3. Interleave the I and Q data, and byte swap if using little endian order.

This step has two sets of code:

Interleaving and byte swapping I and Q data for little endian order

Interleaving I and Q data for big endian order

For more information on byte order, see

“Little Endian and Big Endian (Byte Order)” on page 186

.

Line

Code Description—Saving the I/Q Data to a Text File

12
13
14
15
16
17
18
19

char *ofile = "c:\\temp\\iq.txt";

FILE *outfile = fopen(ofile, "w");

if (outfile==NULL) perror ("Error opening file to write")

;

for(index=0; index<numsamples; index++)

{

fprintf(outfile, "%d, %d\n", idata[index], qdata[index]);

}

fclose(outfile);

Line

Code Description—Saving the I/Q Data to a Text File

12

Set the absolute path of a text file to a character variable. In this example, iq.txt is the file name and *ofile
is the variable name
.

For the file path, some operating systems may not use the drive prefix (‘c:’ in this example), or may require
only a single forward slash (/), or both ("/temp/iq.txt")

13

Open the text file in write format.

14

If the text file does not open, print an error message.

15–18

Create a loop that prints the array of generated I and Q data samples to the text file.

19

Close the text file.