1.2 Specifications

The following specifications are based on a one-year calibration cycle, an operating temperature of 18 to
28ºC (64 to 82ºF) and a relative humidity not exceeding 80%.

Table 1-1 List of Specifications

TUNER POWER METER SPECIFICATIONS

Power Range

10 pW to 10 kW (detector dependent)

Power Scales

(photodetector)

Autoscale

Power Scales

(thermal head)

25 scales: 10nW, 30nW, 100nW, 300nW, 1uW, 3uW, 10uW, 30uW, 100uW,

300uW, 1mW, 3mW, 10mW, 30mW, 100mW, 300mW, 1W, 3W, 10W, 30W,

100W, 300W, 1kW, 3kW, 10kW

Compatible detector
heads

XLP12, all UP Series, PS 1.5k, PS 3k, PS 6k, all HP Series and all PH series

Digital Resolution (XLP-
12 detector)

1 µW

Digital Resolution
(UP/PS thermal
detector)

1 mW

Digital Resolution (HP
detector)

Autoscale

Digital Resolution
(photodetector)

10 pW

Monitor Accuracy

±1 %

5 µV

Response Time
(accelerated)

1

< 1sec

Sampling Frequency

16.7 Hz

Analog out

0-1volt (±1%)

Digital Display

77 x 58 mm LCD

LCD segments

96 (55 segments for needle)

Needle accuracy

0.9%

Display Rate

4 Hz

Dimensions

210 mm(W) x 122 mm (H) x 44 mm (D)

Weight (including
batteries)

0.47 kg

Batteries (included)

4 Alkaline AA batteries

Current consumption
(without backlight)

5mA

Backlight current

60mA (with external 9 VDC power supply only)

Battery life (with
passive thermal

500 hours

1

Varies with detector head.