The SKC BioStage

®

single-stage viable cascade impactors operate on the principle

of inertial impaction and meet NIOSH Method 0800 and 0801 requirements and

ACGIH recommendations for sampling viable microorganisms including bacteria

and fungi. Made of precision-tooled autoclavable aluminum, sampler components

are precision-threaded together for a positive seal to ensure sample integrity. Air is

drawn through the impactor where particles are impacted onto an agar collection

medium for culture, count, and identifi cation. BioStage provides performance

equivalent to the Andersen N-6 and Aerotech A6 (see reference on page 7). Two

models of the BioStage Impactor are available.

Standard BioStage contains a 400-hole jet classifi cation stage and

is operated at 28.3 L/min

BioStage 200 contains a 200-hole jet classifi cation stage and is

operated at 14.15 L/min

Specifi cations

Flow Rate:

Standard BioStage: 28.3 L/min

BioStage

200:

14.15 L/min

Material:

Inlet cone and

base

plate:

Precision-tooled autoclavable

aluminum

O-rings:

Duro 50, BUNA-N (not autoclavable)

Jet Classifi cation Stage: Standard BioStage: 400 holes (0.25-mm hole diameter)

BioStage 200:

200 holes (0.25-mm hole diameter)

Median Cut-point (D

50

): 0.6 μm

Sample Media:

15 x 90 to 100-mm agar plates*

Analysis: Colony

culture

Tubing: 1/4-inch

ID

* Consult a laboratory for information on appropriate agar choice.

† In situations where spore counts are high, positive-hole correction should be used. See Macher,

J., "Positive-hole Correction of Multiple-jet Impactors for Collecting Viable Microorganisms,"

American Industrial Hygiene Journal, 50 (11), 1989, pp. 561-568, available at www.skcinc.com/

pdf/Multiple_Jet_Impactors.pdf

Sample Media

Use appropriate agar in a 15 x 90 to 100-mm agar plate with the Standard BioStage

or the BioStage 200. Plastic or glass agar plates can be used. Verify with your agar

supplier that the plates contain the proper volume of agar to achieve the appropriate

agar height to maintain impactor cut-points. Store agar plates as directed by the

supplier.

Suggested Media

For Bacteria: Tryptic Soy Agar (TSA) or Blood Agar Plates (BAP)
For Fungi:

Potato Dextrose Agar (PDA), Malt Extract Agar (MEA),

Dichloran Glycerol 18 Agar (DG-18), or Corn Meal Agar (CMA)

For information on laboratories that can provide agar plates and analyze

samples, see Laboratories at www.skcinc.com.