3 cont mode – KANOMAX 3887 Handheld Particle Counter User Manual
Page 19

12
4-3 CONT Mode
Setup procedure is shown below.
(Character positions may not correspond to the actual screen.)
Power Up
<MENU>
1.MEASURE MODE
2.DATA PROCESSING
3.OPTIONS
< MEASURE MODE >
1.REPEAT 4.CALC
2.SINGLE 5.REMOTE
3.CONT 6.ISO>4
CONT 0.3/0.5/5.0um
LOC.021 STR:N BEEP:N
FREE REC. 09999 OK
CONT 30B 20:32
WAIT 0.3 0.00E+0/m3
0.5 0.00E+0/m3
5.0 0.00E+0/M3
CONT 30B 20:32
READY 0.3 0.00E+0/m3
0.5 0.00E+0/m3
5.0 0.00E+0/M3
Move the
cursor by
▲▼key.
Press ENTER key
Press ENTER key
Press ENTER key
Press ENTER key
Move the
cursor by
▲▼key.
Measurement Setting Screen
LOC.:
LOC number can be used for identifying room
numbers or measurements. Setting of this item
is not mandatory.
STR:
Used when storing data in the instrument. When
storing the data, select “Y” by using the ▲▼
key. Select “N” if you do not want to store the
data.
BEEP:
Alarm goes off when alarm level is exceeded.
SAMPLE:
Sets sampling time. Setting range is from 10 sec
to 99min 59 sec.
FREE REC.:
Indicates remaining data storage capacity.
Setting of every item is not mandatory.
For example, you can change only the LOC
number and press the START/STOP key, and the
display will switch to a measurement standby
screen.