SKC 225-9013 Preloaded Coated Filter User Manual
SKC Hardware
 
Preloaded Coated Filters
Cat. No. 225-9013
SKC coated fi lters are shipped preloaded in the cassett es with end plugs and 
shrink bands in place. No assembly is required; they are ready to use. 
Method & Chemical: ASTM D5836 for 2,4-TDI and 2,6-TDI. OSHA 42* for 
HDI, 2,4-TDI, and 2,6-TDI. OSHA 47 for MDI
* OSHA Method 42 specifi es a fi lter loading of 0.1 mg of 1-(2-pyridyl) piperazine. OSHA 47 and
ASTM D5836 specify a fi lter loading of 1 mg of 1-(2-Pyridyl) piperazine; however, Coated 
Filters 225-9002 (loading of 2 mg to increase stability and shelf-life of fi lters) and 225-9013 
(loading of 1 mg) are acceptable for all of these sampling methods.
Filter & Coating
†
: Glass fi ber fi lter and cellulose support pad coated with
1-(2-methoxyphenyl) piperazine. Three-piece 37-mm cassett e. Open-face 
sampling confi guration
Prior to Sampling: Store at ≤ 39.2 F (4 C) for up to 3 months.
Sample Stability: Store samples of 2,4-TDI, 2,6-TDI, and HDI at ≤ 39.2 F (4 C) 
for up to 18 days. Refer to the methods above for other compounds.
† Custom order due to very limited shelf-life.
For information on other available
coated fi lters, request SKC Publication 1219.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 Fax: 724-941-1369
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which 
provides SKC's sole liability and the buyer's exclusive remedy. To view the complete 
SKC Limited Warranty and Return Policy
How to Use the SKC Preloaded Coated Filters
1. Select one coated fi lter cassett e for calibrating the fl ow rate. A red plug 
secures the inlet, which is clearly marked “inlet,” and a blue plug secures the 
outlet. Set up the sampling train for calibration with the representative fi lter 
cassett e in line. For “closed-face” sampling, remove the plugs and connect the 
cassett e to the sampling train. For “open-face” sampling, remove the outlet 
plug and the cassett e inlet section then connect the cassett e to the sampling 
train. For details on sett ing up a sampling train, refer to the SKC Sample Setup 
Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired fl ow rate using a fi lm fl owmeter or 
calibrated precision rotameter. For details on calibrating a pump, refer to the 
SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC 
Publication 1163).
3. Replace the cassett e used for calibration with a fresh coated fi lter cassett e 
for sample collection. Att ach the cassett e to a worker’s collar, as close to the 
breathing zone as possible, by using a fi lter cassett e holder (SKC Cat. No. 
225-1). Sample for the specifi ed time interval and record the time.
4. Remove the fi lter cassett e at the end of the specifi ed sampling period and 
replace both end plugs (and the inlet if necessary). Recheck the fl ow rate using 
the same cassett e and fl owmeter used for calibration (in Step 2) to ensure that 
the fl ow rate has not changed by more than 5%.
5. See Sample Stability on reverse side for sample storage.  Analyze samples as soon 
as possible. Appropriately package samples, blanks, and all pertinent data to 
send to a lab for analysis. Ship samples under refrigerated conditions via next 
day air to arrive at the lab on the following business day. Friday shipments 
are not recommended.
Form 40042 Rev 1303