Auto-zero – SKC Limited Haz-Dust IV User Manual

Page 32

Advertising
background image

Auto-Zero,

Continued

2

Insert the Zeroing Filter using the table below.

If Sampling...

Then...

Thoracic

Particulates

(Figure 3.7a)

Insert the zeroing filter into the
Thoracic sampling inlet.

Inhalable

Particulates

(Figure 3.7b)

Insert the zeroing filter (p/n ZA-
202A) into the front of the IOM
front plate refers to figure 3.7b.

Respirable

Particulates

(Figure 3.7c)

Insert the zeroing filter into the
bottom of the GSA-202. GS-Cyclone
adapter.

Figure 3.7a Zeroing filter (p/n ZF-102) being attached to the Thoracic sampling inlet.

Figure 3.7b. Zeroing filter (p/n ZA-202A) being attached to the Inhalable sampling
inlet.

3 - 8

Figure 3.7c. Zeroing filter (p/n ZF-102) being attached to GSAS-202 GS-Cyclone
adapter.

Continued on next page