T R O U B L E S H O O T I N G


ELECTRIC MODELS:

1) If motor fails:

• If motor is extremely cold.

o

Bring to room temperature before starting.

• Ensure respirator and hose are connected when pump is running; failure to do so may cause lock-up.

o

Loosen all six pump endplate bolts slightly. Re-tighten and try to run unit.

• Some motors are thermally protected. An electrical overload or mechanical failure may cause a thermally

protected motor to stop running.
o

Let motor cool down (15-20 min.) and try restarting.


• If the motor fails to start or hums.

o

Pull plug and check for correct electrical current as shown on motor nameplate.

o

Check the wiring instructions on motor case or on thermal plate cover.

• If electrical motor hums but the air pump is not turning.

o

Turn off motor, loosen all six end plate bolts, then re-tighten and start unit.

o

If unsuccessful, remove endplate.

o

Check vanes and rotor for free movement.

o

Broken or chipped vanes should be replaced.

• If rust has formed in pump chamber due to exposure to moisture.

o

Sand down if necessary.

o

Clean chamber with approved solvents.


• If pump becomes noisy.

o

Check for sticking vanes.

o

Flush pump, if necessary replace vanes (see disassembly for instructions)


2) If output air hot or pump is running hot:

• If there is a drastic temperature increase on pump surface or air temperature coming out of the respirator.

o

Check alignment of pump rotor and housing. (see Maintenance section)


• If outlet pressure is too high.

o

Adjust pressure relief valve, while maintaining sufficient air flow to the respirator.


• Check inlet and/or exhaust filter for excessive dirt or dust.

o

Replace if necessary.


• Check airline hose.

o

Loosen hose coils to allow heat to dissipate.

o

Use NIOSH approved airline hose only.

o

Do not use more than 100 feet of airline hose per respirator user.

o

Joining separate lengths of airline hose to complete 100 feet will increase pressure and temperatures.

3) If there is a rise in outlet pressure:

• Check for kinked airline hoses.

• Make sure only NIOSH approved airline hoses are used

• Adjust pressure relief valve, to adjust for any change in number of users, or lengths of airline hose.

o

Check condition of filter elements.

o

Replace if necessary.

8