LF P
ROBE
A
DJUSTMENT
(1000:1)
Connect the probe to a 1 kHz square wave
signal. Adjust compensation trimmer in the
BNC-box for optimum square wave response.
Undershoot
Correct
Overshoot
HF P
ROBE
A
DJUSTMENT
If necessary the probe can be adjusted for high
frequency. Connect the probe to a 1 MHz square
wave signal and adjust the trimme rin order to
receive an optimum square wave response.
Undershoot
Correct
Overshoot
S
TANDARD ACCESSORIES SUPPLIED
WITH THE PROBE
Accessories
Quantities
1. Trimming tool
1
2. Ground lead with a crocodile clip
1
3. Hook
1
H
IGH
V
OLTAGE
P
ROBE
K
IT
P
K
104
Accessories
Quantities
1. Trimming tool
1
2. Ground lead with a crocodile clip
1
3. Hook
1
4. BNC Adapter
1
1
3
4
2
Derating PPE 20 kV
1000
10000
100000
0,01
0,1
1
10
100
Frequency [MHz]
Maximum Impuls Derating - PPE20kV
0
5000
10000
15000
20000
25000
30000
35000
40000
45000
10
100
1000
Duration [msec]
10%
25%
50%
100%
Duty Cycles