Press

Start Index

to open the list sweep start index

parameter.

Press

Stop Index

to open the list sweep stop index

parameter.

Press

Dwell Time

to open the dwell-time-per-step

parameter.

Open the parameter you wish to change, then edit
the current value using the cursor control key or ro-
tary data knob or enter a new value using the key
pad and appropriate terminator key. When you have
finished setting the open parameter, close it by
pressing its menu soft-key or by making another
menu selection.

To go to the List Sweep Trigger menu from this
menu, press

Trigger

. The trigger menu lets you se-

lect a list sweep trigger.

Press

Previous Menu

to return to the main List

Sweep Menu display.

Selecting a
List Sweep
Trigger

There are four modes of sweep triggering in list
sweep frequency mode, each selectable from the
trigger menu. The following is a description of each
mode

q

Auto (Automatic)–The output sweeps be-
tween the specified list start and stop indexes,
dwelling at each list index for the specified
dwell time.

q

External–The output sweep recurs when trig-
gered by an external TTL-compatible clock
pulse to the rear panel

AUX I/O

connector.

q

Single (Trig)–A single output sweep starts
when the trigger key is pressed. If a sweep is
in progress when the key is pressed, it aborts
and resets.

q

Manual(list sweep default trigger mode) The
list index is incremented or decremented by
using the front panel cursor control key. The
list index can also be incremented using an ex-
ternal trigger input. Each trigger increments
the list index by one.