Vanguard ATO-600P User Manual

Page 40

Advertising
background image

ATO-400P / ATO-600P Operating Procedures

40

Table 13.0 Calibration Test Procedures (Continued)

STEP

ACTION

ATO DISPLAY

13-18

Automatic; the ATO will measure the 1900

?

-ohm shunt at 10A test current.

13-19

Automatic; the ATO will measure the 1900

?

-ohm shunt at 100A test current.

13-20

Automatic; Test results will be printed at
the end of the 100A test and the ATO
returns to the START MENU.

Figure 13 Calibration Test Results

CURRENT: 10.0 AMP
BURN IN: 02 SEC
1900

? -OHM TEST

CURRENT: 100.0 AMP
BURN IN: 02 SEC
1900

? -OHM TEST

1. RUN TEST 04/11/05
2. SETUP 12:26:01
3. DIAGNOSTIC

Test Passed

Test Failed

Calibration Test Results Failed

Advertising
This manual is related to the following products: