3 combination remote converters – Yokogawa ADMAG AXF User Manual
Page 4

IM 01E20D01-01E
1. INTRODUCTION
1.3 Combination Remote
Converters
IMPORTANT
• The AXF remote flowtube (size 2.5 (0.1 in.) to
400 mm (16 in.)) should be used in combina-
tion with one of the following converters:
- AXFA11 remote converter
- AXFA14 remote converter
Contact Yokogawa before using it in combina-
tion with converters other than those listed
above.
• The AXF remote flowtube (size 500 (20 in.) to
2600 mm (104 in.)) should be used in combina-
tion with AXFA11 remote converter.
Contact Yokogawa before using it in combina-
tion with converters other than AXFA11.
• In case of TIIS Flameproof type, a remote
flowtube is available for combined use with the
AXFA14 remote converter only.
• If the converter combined with the AXF mag-
netic flowmeter’s remote flowtube is changed
from the AXFA11 to AXFA14 or vice versa, the
meter factor of the remote flowtube must be
readjusted according to its flow calibration.
1-4