IM 61A01A01-01E-A

Desired Display Reading

4mA
Adjustment

20mA
Adjustment
Minimum

20mA
Adjustment
Maximum

SW1

SW2

SW3

SW4

SW5

000

1200

1999

OFF

OFF

OFF

OFF

OFF

000

600

1200

ON

OFF

OFF

OFF

OFF

000

400

600

OFF

OFF

ON

OFF

OFF

000

200

400

ON

ON

ON

OFF

OFF

000

100

200

ON

ON

ON

ON

OFF


(5) Set the Decimal value as follows:

a. If tenths (ex. XXX.X) set SW8 on.
b. If hundredths (ex. XX.XX) set SW7 on.
c. If thousandths (ex. X.XXX) set SW6 on.

(6) The values to check are shown in the table below.


(7) Set the voltage standard to the 0% value. Adjust the Zero Pot (top one) to read the minimum value

needed.

(8) Set the Voltage Standard to the 100% value. Adjust the Span Pot (bottom one) to read the

maximum value needed.

(9) Check all points in the table above and verify unit is within specification.














Suffix –A
4-20mA

Suffix –B
10-50mA

Suffix –C
1-5V

0%

4.0mA

10mA

1.0VDC

25%

8.0mA

20mA

2.0VDC

50%

12mA

30mA

3.0VDC

75%

16mA

40mA

4.0VDC

100%

20mA

50mA

5.0VDC