2. GENERAL SPECIFICATIONS

2.1 Measuring elements

: pH glass electrode

: Silver chloride reference

: Solid platinum electrode

: Pt1000 temperature sensor.

2.2 Wetted parts
Sensor body

: PPS 40GF (Ryton™ with glass filling)

Earthing pin

: Solid Platinum

Measuring sensor

: G-glass

LE glass tube

: AR-glass

O-ring

: Viton

Reference junction

: Porous PTFE

2.3 Functional specifications (at 25°C)
Isothermal point

: pH 7

Reference system

: Ag/AgCl with saturated KCI

Glass impedance

- Dome shape

: 200 M

Ω nominal

- Flat Surface

: 700 M

Ω nominal

Junction resistance

: 1.0 to 15 k

Ω

Liquid outlet

: non-flow double junction

Temperature element

: Pt1000 to IEC 751

Asymmetry potential

: 8 ± 15 mV

Slope

: > 96 % (of theoretical value)

Note: The temperature sensor included in the FU24 is designed for cell compensation

and for indication. It is

NOT designed for process temperature control.

2.4 Dynamic specifications
Response time pH

: t90 < 15 sec. (for 7 to 4 pH step)

Response time temperature

- Dome shape

: t90 < 1 min. (for 10 °C step)

- Flat surface

: t90 < 4 min. (for 10 °C step)

Stabilization time pH

: < 2 min. (for 0.02 pH unit during 10 sec.)

2.5 Operating range
pH

: 0 to 14

ORP

: -1500 to 1500 mV

Temperature

- Dome shape

: -10 to 105 ºC (14 to 221 ºF)

- Flat Surface

: 15 to 105 ºC (59 to 221 ºF)

Pressure

: 0 to 10 bar (0 to 145 PSIG) at 25 ºC

0 to 5 bar (0 to 73 PSIG) at 105 ºC

Conductivity

: > 10 µS/cm

Note: The pH operating range at room temperature is 0-14 pH, but at high temperatures or

range outside 2-12 pH the lifetime will be seriously shortened.

Note: Sensor is suitable for pure water applications