Yokogawa PR10 Retractable Assembly User Manual

Page 14

Advertising
background image

14

IM 12B6K3-01E-E

3.8 Sensor mounting

3.8.1 Mounting a 12mm sensor

in pH12

1. Take the cable out of the box and cut off

the cable tie.

2. Tape the separate wires of the cable

together.

3. Take the fitting out of the box and

remove the sensor holder and Y-cap
adapter.

4. Slide the Y-cap adapter over the cable.
5. Release the pigtail (cable gland) com-

pletely. Do not undo the part in the metal
tube!

6. Lead the sensor cable through the tube

of the fitting, from the side where the
sensor holder has been removed.

7. Take the sensor out of the box.
8. Slide the O-ring (12x1) over the connec-

tor.

9. Connect the sensor to the cable.
10. Slide the sensor holder over the sensor.
11. Screw the Y-cap adapter onto the sensor

holder.

12. Hold the sensor still and turn the metal

tube onto the sensor holder. Don’t rotate
the sensor, but rotate the tube of the
fitting, because the cable can be discon-
nected from the sensor, when rotating it

13. Lead the loose part of the pigtail onto the

cable and screw it onto the fixed part.

14. Remove the tape.

3.8.2 Mounting a 12mm PG13.5

sensor in pH13

.

1. Take the cable out of the box and cut

off the cable tie

2. Tape the separate wires of the cable

together.

3. Take the fitting out of the box and

remove the sensor holder.

4. Release the pigtail (cable gland)

completely. Do not undo the part in the
metal tube!

5. Lead the sensor cable through the tube

of the fitting, from the side where the
sensor holder has been removed.

6. Take the sensor out of the box.
7. Screw the sensor in the sensor holder.
8. Connect the sensor to the cable.
9. Hold the sensor still and turn the metal

tube onto the sensor holder. Don’t
rotate the sensor, but rotate the tube
of the fitting, because the cable can be
disconnected from the sensor, when
rotating it

10. Lead the loose part of the pigtail onto

the cable and screw it onto the fixed
part.

11. Remove the tape.

FIG. 7 Mounting pH sensor with pH12

FIG. 8 Mounting pH sensor with pH13

Y CAP adapter

sensor holder

sensor holder