Yokogawa SC25F User Manual

Page 10

Advertising
background image

IM 12B6J1-41E-E

10

CSA
Certificate no.

: 2516979

Master contract no.

: 182892

IS, Class I Div. 1, GP A, B, C, D T3…T6

Electrical data

: For sensor input circuits (by connector), connected to a

certified intrinsically safe circuit, with the following maximum
values:
Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 µH; Ci = 30 µF
or
Certified intrinsically safe Yokogawa transmitter Model FLXA21
series.

Ambient temperature

: T6 for Tamb. -40 °C to +60 °C

T5 for Tamb. -40 °C to +75 °C
T4 for Tamb. -40 °C to +110 °C
T3 for Tamb. -40 °C to +125 °C

Note: Intrinsically safe when connected as per Control Drawing FF1-K1224QV (see Fig 2)

Control Drawing CSA

The SC25F SENCOM

TM

sensor shall be

installed to a certified intrinsically safe circuit
meeting the entity parameters of the sensor
as shown in the table as maximum values,
or to a certified intrinsically safe Yokogawa
transmitter Model FLXA21 series.

When installing this equipment, follow the
manufacturer’s control drawing. Installation
should be in accordance with Canadian
Electrical Code, Part 1 or CEC, Part 1.

To prevent ignition of
flammable or or combustible
atmospheres, disconnect
power before servicing or
read, understand and adhere
to the manufacturer’s live
maintenance procedures.

81 82 83 84 85 86 87

+ - G

Hazardous

location

Transmitter

IS, Class I Div.1, GP A, B, C, D T3/T4/T5/T6

Ta 125 ºC/110 ºC/75 ºC/60 ºC

WU11 cable

terminals 82-84,

86-87.

Maximum cable

length: 100 m.

Cable diameter:

5.8mm

Sensor

Ui

6.1 V

Ii

230 mA

Pi

1.2 W

Li

4 µH

Ci

30 µF

Maximum values

SC25F - -

120

120mm

225

225 mm

AGP25

General purpose

ALP25

High temp.

chemical

resistant

Fig 2: FF1-K1224QV Control drawing CSA