1

ENGLISH

1. GENERAL

1-1. Dimensions
Installation dimensions and basic construction details of the conductivity cells are shown in
fig.1.

1-2. Type number
The type number of the conductivity cell, shown on the cable, gives coded information on
the cell constant, the materials of construction etc.

1-3. Specifications
General specifications

Wetted parts materials
a. body & electrodes

: Stainless steel AISI 316 or Titanium

b. insulations

: PEEK

c. mounting adapter

: PVDF or Stainless steel

Operating Specifications (Sensor)
Measuring system

: 2-electrode

Maximum pressure

: 10 Bar

Maximum temperature

: 110° C

Temperature response

: < 1 minute for 90% of a step change

Sterilize

: At 135° C

Operating Specifications (Adapters)
See fig.1.0

0

10 20 30 40 50 60 70 80

100

120

140

160

BAR

ºC

2

4

6

8

10

12

SS

PVDF

Figure 1-0. Process/temperature relationship