7. MAINTENANCE

The product starts running immediately when the power is turned

on; however, it needs 10 to 15 minutes of warm-up before it

meets the specified performance.

7.1 Calibration Apparatus

• Two DC voltage/current standard (Yokogawa 7651 or the

equivalent)

• A digital multimeter (Yokogawa 7561 or equivalent)

• A precision resistor of 250 Ω ±0.01%, 1W

Setting tool for adjustment (Refer to

“4.2 Connector for

Communication” in this manual.)

7.2 Calibration Procedure

1. Connect the instruments as shown below.

10

11

3 2 1

4

5

6

7

8

9

11

10

8

L

+

N

GND

1

3

+

+

Ri

Ri

4

6

Voltage and

current

generator

Voltage and

current

generator

2

5

7

9

R

R

DMM

DMM

+

+

Power Supply

External receiving resistor

(Shunt resistor

for current input)

Input-1

Ri: External receiving resistor

for current input

Ro: For current output using

250 Ω precision resistor

Input-2

Output-2

Output-1

2. For the high signal selector: VJSS-H□□-□□□0

Apply the input signal equivalent to -5 to 0% of the input span

to one of two inputs of the converter from a voltage current

source.

For the low signal selector: VJSS-L □□ - □□□ 0

Apply the input signal equivalent to 100 to 105% of the input

span to one of two inputs of the converter from a voltage

current source.

3. Apply the input signal equivalent to 0, 25, 50, 75, and 100%

of the input span to one of two inputs of the converter from a

voltage current source.

4. Check to see the corresponding output voltages are 0, 25, 50,

75, and 100% respectively and within the specified accuracy

rating. “R” is used for current output.

• Use the setting tool (VJ77 Parameter Setting Tool or JHT200

Handy Terminal) to adjust the input/output signals.

Input Adjustment Procedure

(1) Input the value equivalent to 0% value of the input range to

the high/low signal selector.

(2) Call the display item (A: DISPLAY1) to check the input value

in A01: INPUT1.

(3) If the adjustment is necessary, call the adjustment item (P:

ADJUST).

(4) Select P08: IN1 ZERO ADJ to enter the adjustment mode.

Select EXECUTE (adjustment) for adjustment. (If RESET is

selected, the adjusted value is reset to the factory-set default.)

(5) Input the value equivalent to 100% value of input range.

(6) Call the display item (A: DISPLAY1) to check the input value

in A01: INPUT1.

(7) If the adjustment is necessary, call the adjustment item (P:

ADJUST).

(8) Select P09: IN1 SPAN ADJ to enter the adjustment mode.

Select EXECUTE (adjustment) for adjustment. (If RESET is

selected, the adjusted value is reset to the factory-set default.)

• Input-2 can be adjusted by the same operation as the above.

Output Adjustment Procedure

(1) When adjusting 0% value of output, call the adjustment item

(P:ADJUST) to select [P26:OUT1ZERO ADJ].

(2) If there is a positive deviation, correct it by setting a negative

value to offset the deviation. If there is a negative deviation,

correct it by setting a positive value.

*: The 100% value of output-1 and the 0%/100% value of

output-2 can be adjusted by the same operation as the above.

For adjustment using a setting tool, refer to the User’s Manual

for each setting tool and

“6. LIST OF PARAMETERS” in this

manual.

User’s Manual for VJ77 [Document No.: IM 77J01J77-01E];

however, use the VJ77 of version R1.04 or later.

User’s Manual for JHT200 [Document No.: IM JF81-02E]

SAFETY AND EMC STANDARDS

CAUTION

This instrument is for Measurement Category I (CAT.I).

Do not use it for measurements in locations falling

under Measurement Categories II, III, and IV.

Internal Wiring

Outlet

Entrance

Cable

III

T

I

II

IV

Measurement

category

Description

Remarks

I

CAT.I For measurements performed

on circuits not directly connected

to MAINS.

II

CAT.II For measurements performed

on circuits directly connected to

the low-voltage installation.

Appliances,

portable

equipments,

etc.

III

CAT.III For measurements performed in

the building installation.

Distribution

board, circuit

breaker, etc.

IV

CAT.IV For measurements performed

at the source of the low-voltage

installation.

Overhead

wire, cable

systems, etc.

An analog input signal is measurement category I (CAT.I).

Rated transient overvoltage: 1500 V

(Note)

Note

This is a reference safety standard value for

Measurement Category I of IEC/EN/CSA/UL61010-1.

This value is not necessarily a guarantee of instrument

performance.

EMC standards: Complies with EN61326.