10.LIST OF PARAMETERS

No.

Item

Display Remark No.

Item

Display Remark

01 Model MODEL

02 Tag no. TAG NO

03 Self-check result SELF CHK

Display items

A Dispay1 DISPLAY1 B Dsplay2 DISPLAY2

A01 Input value INPUT1 B01 Input value INPUT1

A02 Output value OUTPUT B02 Output value OUTPUT

A07 Alarm-1 status ALM1 STATUS B07 Alarm-1 status ALM1 STATUS

A08 Alarm-2 status ALM2 STATUS B08 Alarm-2 status ALM2 STATUS

A54 Status STATUS *1 B60 Self-check result SELF CHK

A56 Rev. no REV NO

A58 Menu rev. MENU REV

*1: The Status is displayed for service personnel to see history records.

A60 Self-check result SELF CHK

Setting items

D Setting (I/O) SET(I/O) E Setting (alarm output) SET(ALM)

D01 Tag no. 1 TAG NO.1 E03 Alarm-1 setting SET POINT1

D02 Tag no. 2 TAG NO.2 E04 Alarm-2 setting SET POINT2

D03 Comment 1 COMMENT1 E05 Direction of alarm-1 action ALM1 ACTION

D04 Comment 2 COMMENT2 E06 Direction of alarm-2 action ALM2 ACTION

D10 Output range unit UNIT E09 Alarm-1 hysteresis HYSTERESIS1

D14 Low output cut point LOW CUT E10 Alarm-2 hysteresis HYSTERESIS2

D16 Input type INP TYPE E11 Alarm-1 ON delay setting ON DELAY1

D17 Selection of input hard range SELECT RANGE E12 Alarm-2 ON delay setting ON DELAY2

D18 Input resistance IN RESIST E13 Alarm-1 OFF delay setting OFF DELAY1

D22 Input low range INPUT1 L_RNG E14 Alarm-2 OFF delay setting OFF DELAY2

D23 Input high range INPUT1 H_RNG E15 Direction of alarm-1 relay action RL1 ACTION

D24 Output low range OUT L_RNG E16 Direction of alarm-2 relay action RL2 ACTION

D25 Output high range OUT H_RNG E60 Self-check result SELF CHK

D46 Pulse width type PULSE TYPE F Setting (communication) SET(COM)

D47 Pulse width time PULSE WIDTH F01 Communication protocol PROTOCOL

D60 Self-check result SELF CHK F02 Address ADDRESS

F03 Baud rate BAUD RATE

F04 Parity PARITY

There are items not displayed depending on what output-2 is specified. F05 Data length DATA LEN

F06 Stop bit STOP BIT

F07 Decimal point position of input INPUT DEC PT

F60 Self-check result SELF CHK

Adjusting items

Test items

P Adjustment ADJUST Q Test TEST

P02 Zero adjustment of input-1 ZERO ADJ Q04 Forced output (alarm) ALM1 TEST

P03 Span adjustment of input-1 SPAN ADJ1 Q05 Forced output (alarm) ALM2 TEST

P17 Adjustment of external input resistance RESISTOR ADJ Q08 Forced output OUT TEST

P60 Self-check result SELF CHK Q60 Self-check result SELF CHK

11.MAINTENANCE

The product starts running immediately when the power is turned on;
however, it needs 10 to 15 minutes of warm-up before it meets the
specified performance.
For cleaning the instrument, use a soft and dry cloth.

11.1 Calibration Apparatus

● Voltage/current generator (Yokogawa 7651 or the equiva-

lent): 1

● A counter(Yokogawa TC100 or the equivalent) or oscillo-

scope (Yokogawa DL1540 or the equivalent): 1

● A resistor (1k

⍀, 1.6 k⍀): 1 each

● 6V battery: 1

11.2 Calibration Procedure

Connect the instruments as shown in Fig.11.1. First adjust the out-
put-1 signal and then the output-2 signal.
Produce input signals equivalent to 0, 25, 50, 75 and 100% of the in-
put span to the converter from the voltage and current generator.
Then, check that the converter's output signal shows the outputs cor-
responding to 0, 25, 50, 75, and 100% of the input span within the
rated accuracy range. Take the same calibration procedure for the
output-2, and confirm that the frequency of the output 2 is the same
as that of the output 1.
When using a oscilloscope, the wave shaping of output pulse can be
confirmed.
If the output signal is out of the rated accuracy range, adjust the out-
put signal level using the Handy Terminal (JHT200). For adjust-
ment, refer to Instruction Manual for VJ77 Parameters Setting Tools
(IM 77J01J77-01) or for Handy Terminal (IM JF82-01).

11

10

8

L+

N–

GND

5

2

1.6k

1k

+

9

7

1.6k

1k

+

10

11

3

2

1

4

5

6

7

8

9

Input signal

3

1

+

R

R:100

Externally connected for current input

Shunt resistor
(Externally connected for current input)

Voltage and

current

Generator

Calibration for output-1

Calibration for output-2

6V
Battery

6V
Battery

Counter or

Osilloscope

Counter or

Osilloscope

Power supply

Fig. 11.1