LumaSense Technologies INNOVA 1316A-1 User Manual
Page 75
Advertising

Chapter 4
_________________________________________________________________________________________
_____________________________________________________________________________
BE6020-14
1316A-1 Multi Gas Monitor
LumaSense Technologies A/S
Instruction Manual
Page 75 of 98
Tick the gas(ses) to be Span calibrated, and key in concentration value(s), see
figure 4.23. Press the “Send” button to perform the Span calibration. Observe
the “Process State” change to “In Progress” during the Span calibration, see
Figure 4.23 The Span Calibration Window to select the gasses and the
concentrations for Span calibration
The text box in the Span Calibration window now displays Span Calibration ACK.
The Span Calibration is in progress.
Advertising