Automated pressure calibrators additel 761, Specifications – Additel 761 Automated Pressure Calibrator Datasheet User Manual

Page 5

Advertising
background image

Pressure / Process Calibration Equipment

Additel Datasheet 2014 / 2015

Corporate Headquarters

22865 Savi Ranch Parkway Ste F

Yorba Linda, CA 92887, USA

Salt Lake City Office

802 Bamberger Dr, Suite C

American Fork, UT 84003, USA

Phone: 714-998-6899

Fax: 714-998-6999

Email: [email protected]

Phone: 801-756-5570

Fax: 714-998-6999

Email: [email protected]

Automated Pressure Calibrators

Additel 761

Pressure Ranges

Electrical Measurement and Source Accuracy

SPECIFICATIONS

* One year accuracy (including yearly stability) except ADT761-LLP which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.
* * Additel 761 calibrators support 160 series intelligent digital pressure modules that are available for gauge, vacuum and absolute pressure
from -15 psi to 10,000 psi (-1 bar to 700 bar). Accuracy from 0.025%FS includes operation over 14°F to 122°F (-10°C to 50°C), one year
stability and calibration uncertainty. For detailed specification refer to 160 series pressure modules datasheet.

Type

Model

Pressure Range

Pressure Module 1

Pressure Module 2

Range

Accuracy

Range

Accuracy

Low/Differential

Pressure

ADT761-LLP

-25 to 25 mbar

(-10 to 10 inH

2

O)

-2.5 to 2.5 mbar

(-1 to 1 inH

2

O)

0.05%FS

-25 to 25 mbar

(-10 to 10 inH

2

O)

0.05%FS

ADT761-D

-0.95 to 1 bar

(-13.5 to 15 psi)

-25 to 25 mbar

(-10 to 10 inH

2

O)

0.05%FS

-0.95 to 1 bar

(-13.5 to 15 psi)

0.02%FS

Gauge Pressure

ADT761-L

-0.95 to 7 bar

(-13.5 to 100 psi)

-0.95 to 2.5 bar

(-13.5 to 35 psi)

0.02%FS

0 to 7 bar

(0 to 100 psi)

0.02%FS

ADT761-M

-0.90 to 25 bar

(-13 to 375 psi)

-0.9 to 2.5 bar

(-13 to 35 psi)

0.02%FS

0 to 25 bar

(0 to 375 psi)

0.02%FS

ADT761-H

-0.90 to 40 bar

(-13 to 600 psi)

-0.9 to 2.5 bar

(-13 to 35 psi)

0.02%FS

0 to 40 bar

(0 to 600 psi)

0.02%FS

Gauge/Absolute

Pressure

ADT761-LA

-0.95 to 7 bar

(-13.5 to 100 psi)

-0.95 to 2.5 bar

(-13.5 to 35 psi)

0.02%FS

-0.95 to 7 bar

(-13.5 to 100 psi)

0.02%FS

0.05 to 8 bar.a

(1.2 to 115 psi.a)

0.05 to 3.5 bar.a

(1.2 to 50 psi.a)

0.03%FS

0.05 to 8 bar.a

(1.2 to 115 psi.a)

0.025%FS

ADT761-MA

-0.90 to 25 bar

(-13 to 375 psi)

-0.9 to 2.5 bar

(-13 to 35 psi)

0.02%FS

-0.9 to 25 bar

(-13 to 375 psi)

0.02%FS

0.1 to 26 bar.a

(1.7 to 390 psi.a)

0.1 to 3.5 bar.a

(1.7 to 50 psi.a)

0.03%FS

0.1 to 26 bar.a

(1.7 to 390 psi.a)

0.025%FS

ADT761-HA

-0.90 to 40 bar

(-13 to 600 psi)

-0.9 to 2.5 bar

(-13 to 35 psi)

0.02%FS

-0.9 to 40 bar

(-13 to 600 psi)

0.02%FS

0.1 to 41 bar.a

(1.7 to 615 psi.a)

0.1 to 3.5 bar.a

(1.7 to 50 psi.a)

0.03%FS

0.1 to 41 bar.a

(1.7 to 615 psi.a)

0.025%FS

Barometric pressure

ADT761-BP

100 to 1,200 hPa

100 to 1,200 hPa

0.01%FS

NA

NA

Range

Resolution

Accuracy

Voltage Measurement

±30.0000V

0.1mV

±(0.01%RD + 1.5mV)

Current Measurement

±30.0000mA

0.1μA

±(0.01%RD + 1.5μA)

Current Source

0.000 to 22.000mA

1μA

±(0.02%RD + 2.2μA)

Switch Test

If the switch has detective Voltage, the range is from 3V to 24V

DC 24V output

24V ± 0.5V, max: 50mA

Advertising