Gauge Pressure

[1]

P/N

Pressure Range

Media

[2]

Accuracy(%FS)

Burst

Pressure

(psi)

(bar)

V15

-15

-1.0

G

0.025 (0.05)

GP2

2

0.16

G

0.05

GP5

5

0.35

G

0.025 (0.05)

GP10

10

0.7

G

0.025 (0.05)

GP15

15

1.0

G, L

[3]

0.025 (0.05)

GP30

30

2.0

G, L

[3]

0.025 (0.05)

GP50

50

3.5

G, L

0.025 (0.05)

GP100

100

7.0

G, L

0.025 (0.05)

GP300

300

20

G, L

0.025 (0.05)

GP500

500

35

G, L

0.025 (0.05)

GP600

600

40

G, L

0.025 (0.05)

GP1K

1,000

70

G, L

0.025 (0.05)

GP2K

2,000

140

G, L

0.025 (0.05)

GP3K

3,000

200

G, L

0.025 (0.05)

GP5K

5,000

350

G, L

0.025 (0.05)

GP10K

10,000

700

G, L

0.025 (0.05)

GP15K

15,000

1,000

G, L

0.05(0.1)

GP25K

25,000

1,600

G, L

0.1

1.5×

GP30K

30,000

2,000

G, L

0.1

1.5×

GP36K

36,000

2,500

G, L

0.1

1.5×

GP40K

40,000

2,800

G, L

0.1

1.1×

Absolute Pressure

P/N

Pressure Range

Media

[2]

Accuracy(%FS)

Burst

Pressure

(psi)

(bar)

AP5

5

0.35

G

0.1

AP10

10

0.7

G

0.1

AP15

15

1.0

G

0.1

AP30

30

2.0

G

0.1

AP50

50

3.5

G

0.1

AP100

100

7.0

G , L

0.1

AP300

300

20

G , L

0.1

AP500

500

35

G , L

0.1

AP1K

1,000

70

G , L

0.1

AP3K

3,000

200

G , L

0.1

AP5K

5,000

350

G , L

0.1

Differential Pressure

P/N

Pressure Range

Media

Accuracy

(%FS)

Burst

Pressure

Static

Pressure

Range

(inH

2

O)

(mbar)

DP1

±1

±2.5

G

0.05

100×

±10 psi

DP2

±2

±5.0

G

0.05

100×

±10 psi

DP5

±5

±10

G

0.05

50×

±10 psi

DP10

±10

±25

G

0.05

20×

±10 psi

DP20

±20

±50

G

0.05

20×

±10 psi

DP30

±30

±75

G

0.05

20×

±10 psi

DP50

±50

±160

G

0.05

±10 psi

DP150

±150

±350

G

0.025

50 psi

DP300

±300

±700

G

0.025

50 psi

Electrical Measurement Accuracy

Range

Accuracy

Voltage DC

±30.0000V

± (0.01%RD + 1.5 mV)

Current DC

±30.0000mA

± (0.01%RD + 1.5 μA)

DC 24V

24V±0.5V max:50mA Protect at: 120mA

Switch

[1]

Status OPEN/CLOSED

Compound Pressure

P/N

Pressure Range

Media

Accuracy(%FS)

Burst

Pressure

(psi)

(bar)

CP2

±2

±0.16

G

0.05

CP5

±5

±0.35

G

0.025 (0.05)

CP10

±10

±0.7

G

0.025 (0.05)

CP15

±15

±1

G

0.025 (0.05)

CP30

-15 to 30

-1 to 2

G

0.025 (0.05)

CP100 -15 to 100 -1 to 7

G, L

0.025 (0.05)

CP300 -15 to 300 -1 to 20

G, L

0.025 (0.05)

Pressure Ranges

Note: [1]. Sealed gauge pressure for above 1000 psi

[2]. G=Gas, L=Liquid
[3]. 0.025% FS for gas media only

Note: [1] 1V~12V if switch has detective voltage

SPECIFICATIONS

* One year accuracy (including yearly stability) except DP1 and DP2 range

which is 0.05%FS calibration accuracy and 0.05%FS yearly stability.

Advertising