Measurement Accuracy

Measure and Simulate

Standard

Temperature Range (°C)

Accuracy (°C)

Measure (2W/3W)

Measure (4W)

Source

Pt10(385)

IEC 751

-200 to 850

-100 to 200

0.65

0.60

0.65

200 to 600

0.82

0.72

0.82

600 to 850

0.96

0.82

0.96

Pt100(385)

IEC 751

-200 to 850

-100 to 200

0.15

0.1

0.15

200 to 600

0.26

0.16

0.26

600 to 850

0.34

0.20

0.34

Pt100(3916)

JIS 1604

-200 to 850

-100 to 200

0.15

0.1

0.15

200 to 600

0.26

0.16

0.26

600 to 850

0.33

0.20

0.33

Pt100(3926)

Minco Application Aid #18

-200 to 850

-100 to 200

0.15

0.1

0.15

200 to 600

0.26

0.16

0.26

600 to 850

0.33

0.20

0.33

Pt500(385)

IEC 751

-200 to 850

-100 to 200

0.20

0.16

0.36

200 to 600

0.32

0.22

0.54

600 to 850

0.40

0.27

0.67

Pt1000(385)

IEC 751

-200 to 850

-100 to 200

0.1

0.05

0.25

200 to 600

0.2

0.10

0.42

600 to 850

0.27

0.14

0.54

Cu10(427)

Minco Application Aid #18

-100 to 260

-100 to 260

0.61

0.56

0.61

Cu50(428)

GOST 6651-94

-50 to 150

-50 to 150

0.17

0.13

0.17

Cu100(428)

GOST 6651-94

-50 to 150

-50 to 150

0.12

0.09

0.12

Ni120(672)

Edison curve #7

-100 to 260

-100 to 260

0.07

0.05

0.07

Ni100(618)

DIN 43760

-100 to 260

-100 to 260

0.08

0.06

0.08

SPECIFICATIONS

ORDERING INFORMATION

ADT221A

* Additel/Land software can be downloaded for free at www.additel.com

Model Number

Accessories (included)

110V/220V external power adapter

1 pc

Chargeable Li-ion battery

1 pc

Test leads

3 sets(6 pcs)

Short circuit cable

1 sets(2 pcs)

Manual

1 pc

NIST traceable calibration certificate

1 pc

Optional Accessories

Model number

Description

9050

USB to RS232 (DB-9 Male) Adapter

9080

Cable kit (including TC plug, compensation cable,

S,R,B,K,J,T,E,N)

9712

Spare chargeable Li-ion battery for multifunction

calibrator

9816

110V/220V external power adapter for multifunction

calibrator

9906

Carrying case for multifunction

put next to multifunction

9510

Additel/Cal Task management software for multifunction

calibrator

Advertising