Features

•

Range: 0 to 1000 or 0 to 100 NTU

•

Fast and Easy calibration

•

Modular construction allows easy clean-
ing and servicing.

•

EPA 180.1—Krypton filled white light
with 7 year lamp lifetime

•

Alternate infrared light source to meet
ISO 7027 with 7 year lifetime

•

Small sample volume to minimize cali-
bration standard cost

•

Removable sample cuvettes for easy
cleaning and calibrating

•

Bubble rejection system

•

Automatic continuous ultrasonic optical
cleaning system

•

Optics isolated from sample to reduce
the chance of false low readings

•

Selectable 4-20 mA or RS 485-Modbus

Aquametrix Model AM-TBR

Turbidimeter

by

Water Analytics

AquaMetrix

100 School Street

Andover, MA 01810

978-749-9949

Toll free - 855-747-7623

www.WaterAnalytics.net

Applications

•

Municipal Drinking Water

•

Municipal Wastewater

•

Industrial Wastewater

Description

by

Water Analytics

AquaMetrix

New

The AM-TBR builds on a 35 year old heritage of measuring turbidity in wa-
ter and wastewater installations around the world. The instrument comes
configured with a white light source to meet EPA Method 180.1 or with an
infrared source to comply with ISO 7027.
Unlike other incandescent light sources, the Krypton filled bulb provides
up to 7 years of lifetime. Thee IR LED light source also provides 7 years of
trouble free service.
Optical windows are isolated from the water stream to eliminate false read-
ings. To eliminate artificially low reading caused by biofouling an optional
ultrasonic cleaning mechanism automatically
keeps the sample cuvette clean and free of organ-
ics.
The AM-TBR was designed to be cost effective to
purchase and even more cost effective to main-
tain. The turbidimeter is modular to make servic-
ing and cleaning quick and easy. The compact
sample volume of 30 ml means that you will go
through less calibration solution over time. The
small sample size also means a faster response
time.
Optional Windows software allows the user to
graphically display data history, alarms and ana-
lyze multiple networked turbidimeters.
The AM-TBR can use our primary calibration fluid
instead of unstable and hazardous Formazine.