Type-52 [for solvent gases (%lel) – COSA Xentaur SF6 Purity Pro User Manual

Page 26

Advertising
background image

26

TYPE-52 [FOR SOLVENT GASES (%LEL)]

Selectable Measuring Gas and Range

Measuring Gas

(Display Indication)

Measuring

Range

(Minimum

Digit)

AIR CAL

Value

Drift

Amount

Spec. No.

1

Toluene in AIR

(C7H8 in AIR)

0~100 %LEL

(0.2)

0.0 %LEL 1.4 %LEL

SPE- 1434

2

Methyl ethyl ketone

in AIR

(MEK in AIR)

0~100 %LEL

(0.2)

0.0 %LEL 1.5 %LEL

SPE- 1435

3

Ethyl acetate in AIR

(EtAc in AIR)

0~100 %LEL

(0.2)

0.0 %LEL 1.1 %LEL

SPE- 1436

4

Xylene in AIR

(C8H10 in AIR)

0~100 %LEL

(0.2)

0.0 %LEL 1.4 %LEL

SPE- 1437

5

Isopropyl alcohol in

AIR

(IPA in AIR)

0~100 %LEL

(0.2)

0.0 %LEL 1.7 %LEL

SPE- 1438

6

Methyl isobutyl

ketone in AIR

(MIBK in AIR)

0~100 %LEL

(0.2)

0.0 %LEL 1.3 %LEL

SPE- 1439

7

Methanol in AIR

(CH3OH in AIR)

0~100 %LEL

(0.2)

0.0 %LEL 2.0 %LEL

SPE- 1447

8

-----------------

-----------------

-------------- ----------------- -----------------

Interferometer Type: Type 120MM

Indication Accuracy: ± (Indication Value – AIR CAL Value) x 0.03 ± Drift Amount

AIR CAL Value: Concentration indicated on the display when AIR CAL key is pressed.

Drift Amount: The amount of indication drifting when temperature changes ±10ºC.

Special Caution:

4) This instrument is not explosion proof.

5) This instrument measures combustible gas around lower explosive limits in this

specification. Take enough sample for the measurement.

6) Measuring gases in this specification are not extensively vaporized. The gas

concentration may not reach a higher concentration in themeasuring range depending

on a condition of temperature and pressure.