DeFelsko PosiTector 6000 v.4.0 User Manual

Page 12

Advertising
background image

12

Two-point Optimization

(FT, FTS, NT, NTS, FNT, FNTS, FHS, NHS and
Microprobes only)

This method requires taking a reading on
both the uncoated object (

ZERO

ZERO) and on

a known thickness (

2ND PNT

2ND PNT). This

second point should be near the top end
of your application thickness to be
measured.

1. “

ZERO?

ZERO?” the Gage as described in

“Average Zero Adjustment” (p.10).

2. Select the

2ND PNT?

2ND PNT? (Main Menu)

option.

3. Press 9

9 to set the number of readings

used to obtain an average, typically 3
to 9. The greater the variation between
readings, the more readings should be
taken to obtain an average. The
display will show the number of
readings remaining to be taken; i.e.
“

2NDPNT4X

2NDPNT4X”

4. Repeatedly measure the known

thickness (coated part, a plastic shim
or a Coating Thickness Standard) until
the display shows

2PT

2PT=value. The

Gage will display “

WAIT

WAIT” for 2 seconds

between readings to allow the user to
correctly position the probe on the
surface.

5. Lift the probe from the surface and

adjust the displayed reading down 8

8 or

up 9

9 to the known thickness value.

Press 8

8 9

9 simultaneously to accept this

value.