How to measure, Positector 200 - theory of operation – DeFelsko PosiTector 200 v.3.0 Quick Guide User Manual

Page 4

Advertising
background image

3

1. Remove the protective rubber cap from probe.
2. Power-up Gage by pressing the center navigation button.

3. Apply couplant (see pg. 2) to the surface of

the part.

4. Place the probe FLAT on the surface and

press down.

5. Lift the probe when you hear a double BEEP,

- OR - leave probe on the surface in the same location for

continuous measurements.

6. When completely finished, wipe the probe clean of couplant

then return the Gage to the protective pouch. There is no need

to disconnect the probe from the PosiTector during storage.

How to Measure

How to Measure

Couplant

Probe

Coating 1

Substrate

Coating 2

The PosiTector 200

interprets the largest

“echo” within the

selected range as

the coating/coating

or coating/substrate

echo.

PosiTector 200 probes emit a high frequency sound pulse that

travels into the coating via a coupling gel and reflects from ANY

surface that is different in density. Coating thickness readings are

obtained by measuring the time taken for the ultrasonic signal to

propagate from the probe to the coating/substrate interface and

back. The travel time is divided by two and multiplied by the

velocity of sound for the coating to obtain the thickness of the

coating.

PosiTector 200 - Theory of Operation