Meter/battery check, Taking a reading – Delmhorst Instrument Accuscan User Manual

Page 6

Advertising
background image

METER/BATTERY CHECK

hHold the meter in the air and away from any object. Your

fingers must not make contact with the sensor plate while
performing this function.

hPress and hold the check button (3) until the green light

stops blinking.

h

The battery voltage is sufficient if the needle pointer

moves to within the green band (20%) on the wood
scale. Any reading within this green band on the dial is
acceptable. Make sure the sensor pad is not making
contact with anything when checking the meter.

h

If the needle does not read “20” it is likely an indication

of a low battery. If this occurs, change the battery

(9-Volt) immediately. Continued use with a low battery

may cause the meter to go out of calibration. If you have
a fresh battery and the instrument still does not indicate
a proper reading, return it to DELMHORST for service.
See

“Service for Your Meter” section for details.

***A low battery indication is also provided during the read

cycle. If, while taking a reading, the LED blinks, the
battery voltage is below 6%.

TAKING A READING

hSelect the appropriate scale by pressing and releasing either

the Reference Scale or the Wood Scale button.

Hold the

meter in the air and away from any object when making the
scale selection.

hThe meter will turn on, and the LED will flash several

times. This action also performs an auto calibration, which
ensures that meter readings will be stable, repeatable, and
virtually unaffected by environmental conditions.

hWhen the LED stops blinking, the unit is ready to take a

reading.

hPlace the meter on the surface to be tested.

hHold the outer sides of the unit and apply slight forward

pressure on the meter to minimize the amount of air
between the sensor plate and the material being tested.

hIf you want to read continuously, press and HOLD the Stay

On button (1) to keep the unit ON. If you release the Stay On
button, the meter will turn off after 2 minutes 30 seconds.

hTo turn the meter “ON” after 2 minutes 30 seconds, hold it

in the air and away from any object. Press the Stay On
button or appropriate scale button again.

5

Advertising