OPERATING PROCEDURE

1

To check the meter calibration:
Press the

CAL CHK button. The meter should read

80.0 ± 1.0.
Note: Instrument serial numbers from 0100-0584
require pressing both

READ and CAL CHK

buttons simultaneously.

2

To “read” the blocks:

h Push the tinned ends of the lead wires of the block

into two springloaded binding posts on the meter.

h Press the READ button and record the reading on

the Soil Moisture Observation Chart.

h NOTE: Only readings between 5.0 and 100.0 are

valid. Readings outside of this range should be
disregarded.

INTERPRETATION OF METER
READINGS

h Meter readings are interpreted in terms of

“Soil Moisture Tension,” or “Blocks Resistance”
by referring to figure 1 and 1A; in terms of
“available soil moisture” by referring to figure 2.

h The data in the chart below provides guidelines of

“Available Moisture” and “Irrigation Requirements”
for three groups of soil in terms of their texture
make-up.

h Since soils are not strictly divided into three groups

only, but changes from “fine” to “coarse” soil are
gradual, the information below should be used as a
guideline. Consideration should be given to other
factors such as:

Type of crop;
Stage of crop growth and root system;
Climactic conditions;
Structure of sub-soil.