3.1.3 mA Input Calibration

a. Observing proper polarity, connect precision mA source to mA input jacks of

instrument.

b. Set instrument to mA input mode.
c. Set mA A/B switch to “A” position.
d. Adjust source output to 0.000 mA.
e. Adjust R5 on Display board for a display reading of .00 mA.
f. Adjust source output to 25.000 mA
g. Adjust R20 on Analog board for a display reading of 25.00 mA.
h. Repeat steps 3.1.3 through 3.1.3g until no further adjustments are required for correct

display readings.

i. Carefully reassemble instruments
CALIBRATION COMPLETED