GW Instek GOS-6000 Series User Manual
Page 19

GOS-6051/6050/6031/6030 OSCILLOSCOPE
USER MANUAL
⎯ 32 ⎯
shown in figure 5-2. If either probe needs to be adjusted, proceed the step 8.
If either probe does not need to be adjusted, proceed the “Function Check”.
Figure 5-2 Typical Compensation Waveform
8.Adjust the probe by using a small insulated screwdriver. Slowly rotate the
adjustment control until the probe is properly compensated.
GOS-6051/6050/6031/6030 OSCILLOSCOPE
USER MANUAL
⎯ 33 ⎯
5-4.Function Check
When you start to check the operation of your oscilloscope, proceed the
following instruction:
1. Install the ×10 probes onto CH1 and CH2 inputs.
2. Connect the probe tips to the CAL test point of the oscilloscope.
3. Set the oscilloscope controls to display both channels:
VERTICAL:
VOLTS/DIV
0.2V
COUPLING
DC
ALT/CHOP
CHOP
HORIZONTAL:
TIME/DIV
0.5ms
TRIGGER:
MODE
ATO
SOURCE
VERT
COUPLING
AC
SLOPE
The figure 5-3 below illustrates a satisfactory display. The waveform should
be approximately 0.5Vp-p at a frequency of 1kHz that confirms the vertical
and horizontal deflection function of the oscilloscope.
Figure 5-3