Experiments, Insertion loss measurement 1 – GW Instek GRF-3300 Series Student Manual User Manual
Page 22
Advertising

GRF-3300
Manual
3
330
Experiments
Insertion Loss Measurement 1
Objective: Implement a microstrip low pass filter. Refer to Figure 16-12 or R-11
module in the GRF-3300.
p
L
o
L
o
L
1
L
2
L
3
L
4
L
5
L
6
L
7
L
o
L
o
L
p
L
Figure 16-12 Microstrip line low pass filter
9. Prepare the following items.
x
GRF-3300, R-11 (Microstrip Low-Pass Filter) module
x
N-SMA Adaptor x 2
x
RF cable, 75cm x 2
x
Spectrum Analyzer (SA)
10.
Configure the SA as follows.
x
Frequency span: Full Span
x
Resolution bandwidth (RBW): Auto
Span
BW
Full
Span
RBW
Auto/ Manu
RBW
Auto
Step 1:
Step 2:
Figure 16-13 Spectrum analyzer configurations (GSP-830)
11.
Activate and adjust the Tracking Generator (TG) as follows.
x
TG Reference level: 10dBm
x
TG level: 0dBm
x
TG reference value: 0dBm
Advertising