GW Instek USG-Series User Manual User Manual
Page 13

GETTING STARTED
11
Frequency
Settings
Sets the basic start and stop frequency parameter
settings. It also displays the instantaneous
(current) output frequency, as shown below.
Start frequency
Stop frequency
Current frequency
Frequency Status
Bar
When the output is on, the point on the frequency
status bar indicates the instantaneous frequency
that is being output. When the output is off, the
status bar can set the start and stop frequencies.
Start key
Pressing Start will output the selected function.
System
Information
The system information states the serial number,
model and frequency range specifications.
Frequency Offset
Settings
Offsets the frequency by ±50 kHz.
Time Dwell
Settings
The time dwell settings determine how long the
signal will stay (dwell) at each frequency point.
Step Size
The User Step and Point (inversely related) set the
step resolution of the single and continuous sweep
functions in hertz and number of points,
respectively.