Output voltage verification, Default spd-3606 setting – GW Instek SPD-3606 User Manual
Page 21

PERFORMANCE
VERIFICATION
41
Default SPD-3606 setting
The following is the required front and rear panel setting before
running each verification.
Range
60V/3A
60V/3A
30V/6A
30V/6A
60V/3A
30V/6A
•
Output Voltage
•
Tracking Series
•
OVP
•
Output Current
•
Tracking Parallel
OVP setting SW
Normal position
CH3
CH2
CH1
O.V.P. SET
NORMAL
Channel1/3 meter
SW
CH1 position
CH1
CH3
Tracking SW
Independent position, + .
TRAC
KING
INDE
P.
SERI
ES
PARA
LLEL
Channel1/2/3
Voltage knob
Minimum position
VOLTAGE
Channel1/2
Current knob
Minimum position
CURRENT
SPD-3606 User Manual
42
Output Voltage Verification
Check item
•
Minimum output Voltage accuracy
•
Maximum output Voltage accuracy
•
Voltage meter accuracy (Output On)
•
Voltage meter accuracy (Output Off)
Connection
Digital Multimeter
SPD-3606
V
A
Verification step
1. Set the SPD panel according to the Default
setting list, page41.
2. Connect SPD Channel 1 and
Digital Multimeter Voltage
terminal.
CH1
0 60V , 3A
0 30V , 6A
3. Power up SPD and Digital Multimeter.
4. Turn up the SPD Current knob to
the maximum.
CURRENT