Output voltage verification, Default spd-3606 setting – GW Instek SPD-3606 User Manual

Page 21

Advertising
background image

PERFORMANCE

VERIFICATION

41

Default SPD-3606 setting

The following is the required front and rear panel setting before
running each verification.

Range

60V/3A

60V/3A

30V/6A

30V/6A

60V/3A

30V/6A

•

Output Voltage

•

Tracking Series

•

OVP

•

Output Current

•

Tracking Parallel

OVP setting SW

Normal position

CH3

CH2

CH1

O.V.P. SET

NORMAL

Channel1/3 meter
SW

CH1 position

CH1

CH3

Tracking SW

Independent position, + .

TRAC

KING

INDE

P.

SERI

ES

PARA

LLEL

Channel1/2/3
Voltage knob

Minimum position

VOLTAGE

Channel1/2
Current knob

Minimum position

CURRENT

SPD-3606 User Manual

42

Output Voltage Verification

Check item

•

Minimum output Voltage accuracy

•

Maximum output Voltage accuracy

•

Voltage meter accuracy (Output On)

•

Voltage meter accuracy (Output Off)

Connection

Digital Multimeter

SPD-3606

V

A

Verification step

1. Set the SPD panel according to the Default

setting list, page41.

2. Connect SPD Channel 1 and

Digital Multimeter Voltage
terminal.

CH1

0 60V , 3A

0 30V , 6A

3. Power up SPD and Digital Multimeter.

4. Turn up the SPD Current knob to

the maximum.

CURRENT