GW Instek PSM-Series Programming Manual User Manual
Page 29

PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
⎯ 55 ⎯
CURR:TRIG 3.0
TRIG:DEL 3.0
INIT
*TRG
After proceeding above program, the output will be change to 5V/3A
immediately.
z
CALIBRATION COMMANDS
CALibration:PASSword
Function:
Input Password to enter calibration mode.
Syntax:
CALibration:PASSword <password>
The Password for PSM-2010 is 2010.
The Password for PSM-3004 is 3004.
The Password for PSM-6003 is 6003.
Examples:
CALibration:PASSword 2010
for PSM-2010 to enter calibration
mode
CALibration:TYPe
Function:
Select calibration mode.
Syntax:
CALibration:TYPe {0|1|2|3}
0:Voltage, 1:OVP , 2:Current 3:Ocp
PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
⎯ 56 ⎯
CALibration:VOLTage:MINimum
CALibration:VOLTage:MIDDle
CALibration:VOLTage:MAXimum
Function:
Input the voltage value separately.
Syntax:
CALibration:VOLTage:MINimum <Decimal numbers>
CALibration:VOLTage:MIDDle <Decimal numbers>
CALibration:VOLTage:MAXimum <Decimal numbers>
CALibration:VOLTage:PROTection
Function:
Auto calibrate OVP.
Syntax:
CALibration:VOLTage:PROTection
CALibration:CURRent:MINimum
CALibration:CURRent:MIDDle
CALibration:CURRent:MAXimum
Function:
Input current value separately.