1. Insert the red test lead into the V

Ω

mA terminal and the black test lead into the

COM terminal.
2. Set the rotary switch to an appropriate measurement position in

Ω

range.

3. Connect the test leads across with the object being measured.
The measured value shows on the display.
Note

l

The test leads can add 0.1

Ω

to 0.3

Ω

of error to resistance measurement. To obtain

precision readings in low-resistance measurement, that is the range of 200

Ω

,

short-circuit the input terminals beforehand and record the reading obtained

(called this reading as X). (X) Is the additional resistance from the test lead.

Then use the equation:
measured resistance value (Y) (X) = precision readings of resistance.

l

For high-resistance measurement (>1M

Ω

), it is normal taking several seconds

to obtain a stable reading.

l

When resistance measurement has been completed, disconnect the connection

Measurement Operation (9)